CALIBRATING THE RECORDER

Factory Calibration Information
The recorder is factory calibrated during final assembly by calibrating the sensor (shipped with the recorder) to known temperature and pH standards. This technique avoids tracking and positioning errors due to temperature changes and mechanical shock. However, if the pH readings drift or appear out of calibration, follow Section 5.2 to recalibrate your recorder.

Changing Calibration Codes
To change the calibration codes:
1. Open the control panel door (unlock it, if necessary).

2.
Turn the power switch to the off (ôOö) position.

3.
Press and hold down the ôSETö button while turning on the power switch. This button is one of three Alarm buttons shown in Figure 4-5.
After you hear an audible beep, you will see the following display:
Then ôCHö appears on the display.

4. At this time, release the ôSETö button.

A 2-character temperature calibration code "AC" should now appear in the display with the first character blinking. Refer to Figure 5-2.

First Display
Figure 5-1. First Display
Temp Code - First Character Blinking
Figure 5-2
Temperature Code - First Character Blinking

5a. If the first character is correct,
press the ôSETö button and go to step 6.

5b. If the first character is not correct,
press the ôHIGHö button to increase the value of the character or the ôLOWö button to decrease the value of the character. When this character is correct, press the ôSETö button.

6. The second character should now be blinking. Refer to Figure 5-3. If it is not, press the ôSETö button until it does.


Temperature Code - Second Character Blinking
Figure 5-3. Temperature Code - Second Character Blinking
Blinking Temperature Code
Figure 5-4. Blinking Temperature Code

7a. If the second character of the temperature calibration code is correct,
press the ôSETö button and go to step 8.

7b. If the second character is not correct,
press the ôHIGHö button to increase the value of the character or the ôLOWö button to decrease the value of the character. When this character is correct, press the ôSETö button.

8. A single character pH calibration code "6" should be blinking in the display. Refer to Figure 5-4. If it is not, press the ôSETö button again.

9a. If the temperature code is correct,
press the ôSETö button and go to step 10.

9b. If the temperature code is not correct,
press the ôHIGHö button to increase the value of the character or the ôLOWö button to decrease the value of the character. When this character is correct, press the ôSETö button to lock in the values.
At this time, you see the temperature and pH codes on the display as you entered them. This way, you can verify that you entered them correctly.

10. The unit should now automatically go through the power up sequence as described in Chapter 3. If it does not, hold down the ôSETö button until it does.

While the unit goes through the power up sequence, observe the display to see that the calibration values are what they should be and that they have been entered correctly.

11. If either the temperature or pH code is incorrect,
repeat steps 1-10. After you are all done with the changes, lock the control panel door, if necessary. Refer to Section 4.3 for details.

Calibrating pH Readings
Calibration Setup
Figure 5-5. Calibration Setup
In order to calibrate the pH readings you need 7.00 and 4.00 buffer solutions. Start by turning the unit ON and observing the Display power up routine. After you hear a beep, you will see the display "LO BAT.-188", than the temperature code "Ac", followed by pH code "6".

Put the display switch on pH.

Immerse the sensor in 7.00 buffer solution like in figure 5-5. then wait one minute for the pens to stop moving. Observe the display: it should read 7.0. If it does not, turn the calibration switch to "CAL" and open the door.

When the pens stop moving, turn the "CAL" adjustment to display 7.0, than put the sensor in 4.00 buffer solution and wait one minute. If the display doesn't read 4.0, turn the "Slope" adjustment to bring it to 4.0.

Repeat these steps until the readings are correct in both 7.00 and 4.00 buffer solutions. Now the recorder is calibrated.

Close the door and put the calibration switch on "NORMAL".
Help Price List Specifications Print Manual Manuals Home
 Home Home  Table of Contents TOC  Back Back  Next Next